Recording Hand-Held Laparoscopic Instrument Motion in the Operating Room: Magnetometer-Free Fusion of Inertial, Range and Visual Sensing
Most minimally invasive procedures are performed with hand-held instruments, yet only the endoscopic video is kept and the instrument motion is lost. A clip-on logger fuses a six-axis IMU and a time-of-flight rangefinder with a markerless camera in an error-state Kalman filter under the trocar’s remote centre of motion, with the shaft silhouette supplying heading in place of the magnetometer — whose field changed with tool pose and between sessions in our own measurements. Against a Franka Research 3, and without alignment to it, displacement error was 1.21 mm RMS over 300 trials and relative rotation 0.34° RMS over 180, with 1.22–3.04 mm absolute tip error on continuous trajectories tracked and displayed live.